TM 11-6130-413-12
tiate operational cycle. Unit will discharge battery
CAUTION
continuously; cell voltage must be monitored by opera-
Before connecting or disconnecting a battery,
tor. If desired, set 30V-3V switch (s3) to 3V position,
make sure the charger/analyzer is in RESET.
connect test leads to (+) and (-) 3V terminals (J4, J3)
Do not connect or disconnect a battery when
and monitor cell voltage on VOLTMETER (M4). Short
the RUN light is on.
individual cells in accordance with the battery manual
NOTE
when the monitored cell voltage drops to 0.5 volts; this
The capacity test (initial discharge) and
will ensure that each cell is at zero volts.
(3) Proceed to next phase when cell voltages are at
discharge racks are used.
a. Phase 1Capacity Test (Initial Discharge).
zero (deep discharge).
c. Phase 3, 4, 5-Initial Charge, Capacity Test,
Proceed as follows:
Final Charge (Auto Cycle). This is a continuation of
(1) Set ON-OFF switch (CBl) to OFF position.
Phase 2; proceed as follows:
(2) Set CELL SELECT to match battery to be
(1) Set 30V-3V switch (S3) to 30V position and
charged.
(3) Set AUTO-CHARGE No. 1 timer (M3) fully
disconnect any test leads used in phase 2.
ccw then back to 1 HR position.
(2) Set AUTO-CHARGE No. 1 timer (M3) to
START position.
(4) Set AUTO/MANUAL DISCHARGE timer (M2)
(3) Set AUTO/MANUAL DISCHARGE timer (M2)
to START position.
to START position.
(5) Set AUTO/MAN CHARGE No. 2 timer (MI)
fully ccw then back to 1 HR position.
(4) Set AUTO/MAN CHARGE No. 2 timer (MI) to
START position.
(6) Set ON-OFF switch (CB1) to ON position.
(5) Set FUNCTION SELECT switch (S4) to AUTO
(7) press POWER RESET switch (S5) momentar-
ily.
position.
(8) Set FUNCTION SELECT switch (S4) to AUTO
(6) Set DISCHARGE AMP-FIR (S7), CHARGE
position.
AMP-HR(R-16), DISCHARGE AMP (R-15) and CELL
(9) Set 30V-3V switch (S3) to 30V position.
SELECT (S6) to appropriate positions, according to
(10) Set DISCHARGE AMP-HR switch (S7) to ap-
battery type.
(7) Press START switch (S2) momentarily to ini-
propriate position for l-hour rate.
(11) Set CHARGE AMP-HR control (R16) to ap-
tiate automatic cycle and observe that the following
propriate position for l-hour rate. Note that battery
actions occur:
(a) RUN indicator (DS6) illuminates.
should trickle-charge during the final 5 to 10 minutes
of the charge cycle; adjust control accordingly.
(b) Automatic charge indicator (DS5) illum-
(12) Set DISCHARGE AMP control (R15) to ap-
inates.
(c) CURRENT meter (M5) reads full scale for
propriate position. Use ammeter for precise setting.
(13) If fail light comes on, clear it by moving
less than a second, falls back to 0, then rises again to
AUTO-CHARGE No. 1 tier cw then back to 1 HR
full scale at next cycle. (Note that unit emits an
position.
audible deep vibration each charge cycle, in synchron-
(14) Press START switch (S2) momentarily to ini-
ization with the meter movement).
tiate operational cycle and observe that the following
(d) VOLTAGE meter (M4) reads 26-28 volts af-
actions occure.
ter 40 to 50 minutes and unit continues to charge bat-
(a) RUN indicator (amber, DS6) illuminates,
tery, at a lower rate, for remainder of cycle.
(b) Discharge indicator (amber, DS3) illum-
(CURRENT meter (M5) still pulses, but for shorter
inates.
time periods during each interval; voltage meter is
(c) Battery VOLTAGE (M4) and CURRENT
read during non-active part of pulse cycle.
(M5) meters display battery condition.
(e) Monitor cell VOLTAGE during the last 15
(d) PASS (DS2) or FAIL (DS4) indicator illum-
minutes of the charge cycle.
inate at EOD. PASS indicates that battery discharged
(f) After a 60-minute period, AUTO CHARGE
for at least 1 hour before reaching EOD.
NO. 1 timer (M3) stops, the associated run indicator
(15) Monitor cell VOLTAGE during the last 15
(DS5) extinguishes, AUTO/MANUAL DISCHARGE
minutes of discharge.
timer (M2) begins operation and the associated run
indicator (DS3) illuminates.
is a continuation of Phase 1; proceed as follows:
(g) Battery discharge continues at a constant
(1) Set FUNCTION SELECT switch (S4) to MAN-
rate until EOD is reached (0.95 volts/cell).
(h) If EOD is reached after the 60-minute peri-
UAL DISCHARGE position, and the DISCHARGE
timer to START position.
od, PASS indicator (DS2) illuminates, AUTO/MAN-
(2) Press START switch (S2) momentarily to ini-
UAL DISCHARGE time (M2) stops (providing an indi-
Change 1
3-3