TM 11-6625-2631-14
Item
Corrective action
No.
Probable cause
Malfunction
1
a. Zero meter. using zero adjustment.
a. Movement not properly zeroed:
Meter does not return to zero or rests
b. Replace test set.
b. Sticking movement, broken
below zero when batteries are not
pivots. bent pointer.
connected.
2
a. Replace test set.
c. Sticking movement. broken pivots,
Meter indication is irregular.
bent pointer.
b. Loose or dirty connector contacts.
b. Clean or tighten contacts as
required.
3
a. Replace test set.
a. Defective meter.
No meter indication regardless which
b. Check wiring against schematic
b. Defective wiring.
type of battery is tested.
4
a. With AN / USM-223, check
a. Defective resistor in incremental
No meter indication or faulty meter
values of RI, R2, R3 against
indication only when tasting K308,
K308A, URC-10 batteries.
Replace defective resistors.
b. Defective Zener diode.
b. With AN/ USM-223 connected
across CR 1 [positive lead to
cathode: negative lead to anode ).
and any one of the three tested
batteries connected. check that the
voltage drop across CR 1 does not
exceed 9.1 volts 5%. Replace
CR 1 if incorrect voltage is ob-
tained.
c. Check wiring and circuit board
c. Defective wiring.
Repair or replace, as necessary.
5
a. With AN/ USM-223, check value
a. Defective resistor in incremental
No meter indication or faulty meter
of R4, R5, R6 against schematic
indication only when testing
AN/ URT-21, AN/ PRC-63,
defective resistors.
AN/ URT-27, or AN/ PRC-90
b. Defective Zener diode.
b. With AN/ USM-223 connected
batteries.
across CR 2 [positive lead to
cathode; negative lead to anode ),
and any one of the four tested
batteries connected, check that the
voltage drop across CR 2 does not
exceed 3.8 volts 5% . Replace
CR2 if incorrect voltage is ob-
tained.
c. Check wiring and circuit board
c. Defective wiring.
terminations against schematic
replace, as necessary.
a. With AN/ USM-223, check value
6
No meter indication. or faulty meter
circuit.
of R7, R8 against schematic
indication only when testing SDU-
5 / E battery.
defective resistors.
b. Check wiring and circuit board
b. Defective wiring.
terminations against schematic
replace, as necessary.
Check connector and wiring for
Defective test connector or related
No meter indication, or faulty meter
wiring.
indcation when only one type of
necessary.
battery is tested.
6-3